A published “50 mW/cm²” describes what reaches a sensor under defined conditions — often at the emitter aperture or 1 cm — not what couples into tissue or reaches mitochondrial photoacceptors. Lambertian LED output falls steeply with distance and angle, so a spec-sheet number is a best case.
Irradiance measured at a probe tip can misrepresent the treatment surface by orders of magnitude (Khan & Arany, 2016), and FDA’s 2023 draft guidance expects manufacturers to characterize wavelength, radiant power, irradiance, fluence, output mode and measurement method — not irradiance alone.